Agilent Technologies B1500A Service Manual Page 472

  • Download
  • Add to my manuals
  • Print
  • Page
    / 598
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 471
4-156 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
PWI
Syntax PWDCV chnum,mode,base,start,stop,step
Parameters chnum : MFCMU source channel number. Integer expression. 1 to 10 or 101 to
1001. See Table 4-1 on page 4-16.
mode : Sweep mode. Integer expression. 1 or 3.
1: Linear sweep (single stair, start to stop.)
3: Linear sweep (double stair, start to stop to start.)
base, start,
stop : Pulse base, start or stop voltage (in V). Numeric expression.
0 (initial setting) to ± 100 V.
With the SCUU, the source module is automatically selected by the
setting value. The MFCMU is used if the base, start, and stop values
are below ± 25 V (setting resolution: 0.001 V), or the SMU is used if
they are greater than ± 25 V (setting resolution: 0.005 V).
The SMU will operate with the 100 V limited auto ranging and 20 mA
compliance settings.
step : Number of steps for the pulsed sweep. Numeric expression. 1 to 1001.
Example
Statements
OUTPUT @B1500;"PTDCV 1,0.01,0.02"
OUTPUT @B1500;"PWDCV 9,1,0,-5,5,101"
PWI
The PWI command specifies the pulsed current sweep source and its parameters.
This command clears the settings of the PWV, WSV and WSI commands. The
settings specified by this command are cleared by the PWV command.
To set the timing of output pulse and measurement, use the PT and AIT2 commands.
Syntax PWI chnum,mode,range,base,start,stop,step[,Vcomp[,Pcomp]
]
Parameters chnum : SMU pulsed sweep source channel number. Integer expression. 1 to 10
or 101 to 1001. See Table 4-1 on page 4-16.
mode : Sweep mode. Integer expression. 1 to 4.
1: Linear sweep (single stair, start to stop.)
2: Log sweep (single stair, start to stop.)