Agilent B1500A/B1505A Programming Guide, Edition 11 2-79
Remote Mode Functions
Trigger Function
Table 2-8 Type of Trigger Output
Type Timing of Trigger Output by B1500
Command
a
a. t selects the trigger output terminal, Ext Trig Out or a digital I/O.
p selects positive or negative logic. m selects edge or gate trigger.
1 When the measurement specified by the MM command
is completed.
TGP t,2,p,1
TGXO m
TM3
2 Available for the
staircase sweep, multi channel sweep,
pulsed spot, pulsed sweep, staircase sweep with pulsed bias,
multi channel pulsed spot, multi channel pulsed sweep, and
CV(DC)/CV(AC)/C-f sweep measurements.
For the staircase sweep, multi channel sweep, and
CV(DC)/CV(AC)/C-f sweep measurements: when the
source trigger delay time elapses after the sweep step
output setup or pulse output setup is completed.
For the pulsed spot, pulsed sweep, staircase sweep with pulsed
bias, multi channel pulsed spot, multi channel pulsed sweep
measurements: at 10 μs before starting the pulse output.
TGP t,2,p,2
TGSO m
3 Available for the staircase sweep, multi channel sweep,
and CV(DC)/CV(AC)/C-f sweep measurements.
When the measurement is completed at each sweep
step.
TGP t,2,p,3
TGMO m