Agilent B1500A/B1505A Programming Guide, Edition 11 2-17
Remote Mode Functions
Measurement Modes
2. The pulsed sweep source is set by the MCPNT and MCPWNX commands with
the source identification number N (N=1 to 10).
3. The staircase sweep source is set by the WM and WNX commands with the
source identification number N (N=1 to 10).
4. The pulsed bias source is set by the MCPNT and MCPNX commands with the
source identification number N (N=1 to 10).
5. The DC bias output is started by the DV/DI command.
6. The source output is started by a trigger, such as the XE command. Then the
staircase sweep sources start output in the order specified by the N value, after
that the pulsed sweep sources and the pulsed bias sources start output
simultaneously.
7. After the hold time, all pulse sources start the pulse output simultaneously.
For HR/MP/HPSMU, available delay time value is 0. Also, the pulse width
value must be the same. If a different value is entered, the longest value is set.
8. After the measurement delay time, the measurement channels perform
measurements in parallel. Measurement time is set by the AIT 2 command. This
setting is effective for all measurement channels.
9. The B1500 repeats the step output, the pulse output, and 8 for all sweep steps.
10. After the sweep measurement, the pulsed source forces the pulse base value, and
the sweep source forces the start or stop value, as specified by the WM
command, and keeps it.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the output to 0 V.
NOTE The MCPNT command sets the delay time and pulse width. The MCPWNX
command sets the pulsed sweep output. The base, start, and stop values must have
the same polarity for current pulse or log sweep. The WM command sets the
automatic abort function and the output after measurement.
The MCPNX command sets the pulsed bias output. The base and peak values must
have the same polarity for the current pulse.
The WNX command sets the staircase sweep output. The start and stop values must
have the same polarity for log sweep.