Agilent B1500A/B1505A Programming Guide, Edition 11 2-47
Remote Mode Functions
Automatic Abort Function
Exceptions:
For the sampling measurement, the SPGU output value can be set by the MSP
command, not the MSC command.
For the C-t sampling measurement, the MFCMU output value can be set by the
MDCV command, not the MSC command.
This function is not effective for the pulsed sweep measurement and the pulsed
sweep CV measurement.
NOTE Even if the source output value is specified, the source forces the start value or the
base value if the output is stopped by the automatic abort function, the power
compliance, or the AB command.