Agilent B1500A/B1505A Programming Guide, Edition 11 4-167
Command Reference
RI
This command should only be used for servicing the B1500.
Syntax RCV [slot]
Parameters slot : Slot number where the failed module to enable has been installed. 1 to
10. Or 0 or 11. Integer expression.
0: All failed modules. Default setting.
11: Mainframe.
If slot specifies the slot that installs no module, this command causes an error.
Example
Statements
OUTPUT @B1500;"RCV 1"
OUTPUT @B1500;"*OPC?"
ENTER @B1500;A
Remarks When a SMU connected to SCUU is recovered, the MFCMU is also recovered. The
SCUU cannot be controlled if a failed module is connected.
RI
The RI command specifies the current measurement range or ranging type. In the
initial setting, the auto ranging is set. The range changing occurs immediately after
the trigger (that is, during the measurements). Current measurement channel can be
decided by the CMM command setting and the channel output mode (voltage or
current).
For the high speed spot measurement, use the TI/TTI command.
The range setting is cleared by the CL, CA, IN, *TST?, *RST or a device clear (HP
BASIC CLEAR) command.
Syntax RI chnum,range
Parameters chnum : SMU current measurement channel number. Integer expression. 1 to 10
or 101 to 1001. See Table 4-1 on page 4-16.
range : Measurement range or ranging type. Integer expression. See Table 4-3
on page 4-19.
Example
Statements
OUTPUT @B1500;"RI 2,-20"
RM
This command specifies the auto range operation for the current measurement.