4-198 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
TGPC
TGPC
The TGPC command clears the trigger setting of the specified ports.
Syntax TGPC [port[,port...[,port]...]]
A maximum of 18 ports can be set. If no port is specified, the TGPC command
clears the setting of all ports; Ext Trig In, Ext Trig Out, and digital I/O ports 1 to 16.
Parameters port : Trigger port number. Integer expression. -1, -2, or 1 to 16.
-1: Ext Trig In terminal.
-2: Ext Trig Out terminal.
1 to 16: Port 1 to 16 of the digital I/O terminal.
Remarks The TGPC command sets the trigger ports as shown below.
Ext Trig In Same as after TGP -1,1,2,0 command execution.
Ext Trig Out Same as after TGP -2,2,2,0 command execution.
Digital I/O Ports No trigger function is available. The ERS? and ERC commands
are available for the port control.
This is not same as the condition set by the *RST command that sets the ports as
shown below.
Ext Trig In Same as after TGP -1,1,2,1 command execution.
Ext Trig Out Same as after TGP -2,2,2,1 command execution.
Digital I/O Ports No trigger function is available. The ERS? and ERC commands
are available for the port control.
Example
Statements
OUTPUT @B1500;"TGPC -1,-2,1,2"
See Also “TGP”