1-12 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Basics
Getting Started
To Force Voltage/Current
The commands listed in Table 1-3 is used to force voltage or current. These
commands start to force the voltage or current immediately when the command is
executed. They can be used regardless of the measurement mode.
See Table 1-2 on page 1-11 for the commands available for each measurement
mode. The commands just set the source channel condition, and the source channel
starts the output by the start trigger, such as the XE command. For more details of
the commands, see Chapter 4, “Command Reference.”
Table 1-3 Voltage/Current Output Commands
Example OUTPUT @B1500;"DV 1,0,5"
This example just forces 5 V using channel 1 (the module installed in slot 1 of the
B1500) with auto ranging.
Command Description
DV Applies DC voltage from SMU immediately.
DI Applies DC current from SMU immediately.
FC/ACV Applies AC voltage from CMU immediately.
DCV Applies DC bias from CMU immediately.
TDV Applies DC voltage from SMU, and returns the time data.
TDI Applies DC current from SMU, and returns the time data.
FC/TACV Applies AC voltage from CMU, and returns the time data.
TDCV Applies DC bias from CMU, and returns the time data.