Agilent Technologies B1500A Service Manual Page 414

  • Download
  • Add to my manuals
  • Print
  • Page
    / 598
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 413
4-98 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
ERPFUHCA?
Example
Statements
OUTPUT @B1500;"ERPFUHCA 3,4"
ERPFUHCA?
This query command returns the slot numbers for the MC/HCSMU modules
connected to the V Control and I Control inputs of Agilent N1265A ultra high
current expander/fixture.
Syntax ERPFUHCA?
Query Response vsmu,ismu<CR/LF^EOI>
vsmu : Slot number for the MC/HCSMU connected to the V Control input.
ismu : Slot number for the MC/HCSMU connected to the I Control input.
Example
Statements
OUTPUT @B1500;"ERPFUHCA?"
ENTER @B1500;vsmuUHC,ismuUHC
ERPFUHCCAL?
This query command performs the self-calibration of Agilent N1265A ultra high
current expander/fixture, and returns the execution results. After this command, read
the result soon.
Execution
Conditions
The ERPFUHCA command has been executed before this command.
Syntax ERPFUHCCAL?
Query Response result<CR/LF^EOI>
0: Pass
1: Fail
Example
Statements
OUTPUT @B1500;"ERPFUHCCAL?"
ENTER @B1500;result
ERPFUHCMAX?
This query command returns the available maximum current for the
currently-connected Agilent N1265A ultra high current expander/fixture.
Page view 413