4-92 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
ERPFDA?
Effective slot numbers are 1 to 10. Integer expression. Use the smaller slot number
for the module which occupies two slots, such as HVSMU and HPSMU.
Enter 0 to cancel the assignment.
Example
Statements
OUTPUT @B1500;"ERPFDA 6,0"
ERPFDA?
This query command returns the slot numbers for the measurement resources
connected to the Selector Input of Agilent N1265A ultra high current
expander/fixture.
Syntax ERPFDA?
Query Response hvsmu,smu<CR/LF^EOI>
hvsmu : Slot number for the HVSMU connected to the Selector Input HVSMU
port directly or via the N1266A expander.
smu : Slot number for the MP/HPSMU connected to the Selector Input SMU
port.
Example
Statements
OUTPUT @B1500;"ERPFDA?"
ENTER @B1500;hvSelUHC,smuSelUHC
ERPFDP
This command sets the selector input-to-output connection path of Agilent N1265A
ultra high current expander/fixture.
Execution
Conditions
No measurement resource may be in the high voltage state (forcing more than the
allowable voltage for the interlock open condition, or voltage compliance set to
more than it).
Digital I/O port must be set to the N1265A control mode using the ERMOD 4
command.
Syntax ERPFDP state
Parameters state : Selector input-to-output connection path. Integer expression. 0, 1, 2, 3,
or 4.
0: Open, no resource is connected, initial setting.