LSV
Parameters hold : Hold time (in seconds) that is the wait time after starting the search
measurement and before starting the delay time for the first search
point. Numeric expression.
0 to 655.35 sec. 0.01 sec resolution.
delay : Delay time (in seconds) that is the wait time after starting to force a step
output value and before starting a step measurement. Numeric
expression.
0 to 65.535 sec. 0.0001 sec resolution.
Example
Statements
OUTPUT @B1500;"LSTM 5,0.1"
LSV
The LSV command sets the voltage search source for the linear search measurement
(MM14). After search stops, the search channel forces the value specified by the
LSM command.
This command clears the LSI, LSSI, and LSSV command settings.
This command setting is cleared by the LSI command.
If the output voltage is greater than the allowable voltage for the interlock open
condition, the interlock circuit must be shorted.
Syntax LSV chnum,range,start,stop,step[,Icomp]
Parameters chnum : SMU search source channel number. Integer expression. 1 to 10 or 101
to 1001. See Table 4-1 on page 4-16.
range : Output ranging type. Integer expression. The output range will be set to
the minimum range that covers both start and stop values. Range
changing may cause 0 V output in a moment. For the limited auto
ranging, the instrument never uses the range less than the specified
range. See Table 4-4 on page 4-21.
start, stop : Search start or stop voltage (in V). Numeric expression. See Table 4-7
on page 4-24, Table 4-9 on page 4-26, Table 4-12 on page 4-27, or
Table 4-15 on page 4-28 for each measurement resource type. The start
and stop parameters must have different values.
step : Step voltage (in V). Numeric expression.
If start < stop, step must be positive, and if start > stop, step must be
negative. Maximum number of search steps is 1001.