4-52 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
BSM
2. The source channel forces the Stop value, and the monitor channel executes a
measurement.
If the search target value is out of the range between the measured value at the
Start value and the measured value at the Stop value, the search stops.
3. The source channel forces the Stop-D/2 value (or Stop+D/2 if Start>Stop), and
the monitor channel executes a measurement.
If the search stop condition is not satisfied, the measured data is used to decide
the direction (+ or –) of the next output change. The value of the change is
always half of the previous change.
4. Repeats the output change and measurement until the search stop condition is
satisfied.
For information on the search stop condition, see “BGI” or “BGV”. If the output
change value is less than the setting resolution, the search stops.
Figure 4-2 Binary Search Source Output Control Mode
Start
Stop
D = | Stop - Start |
+D/4
-D/2
-D/8
-D/16
(1) Normal output
Search stopped
+D/32
+D/64
: Measurement
Start
Stop
+D/4
+D/2
-D/8
-D/16
(2) Cautious output
Search stopped
+D/32
+D/64
Time
Voltage or current
Hold time
Delay time
Hold time
Delay time
Delay time
Delay time
Trigger
Trigger
BSM, BST, and
BSV or BSI
BSM, BST, and
BSV or BSI