4-114 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
*LRN?
*LRN?
The *LRN? (learn) query command returns the B1500 command parameter settings.
Syntax *LRN? type
Example
Statements
DIM A$[200]
OUTPUT @B1500;"*LRN? 1"
ENTER @B1500;A$
Parameters and
Query Response
type : This parameter selects the type of query response. Available values are
0 to 110, but some numbers are not used. See below. Integer expression.
A description and the query response of each type is described below.
0 : Returns the output switch ON/OFF status:
CN[chnum[,chnum . . . [,chnum] . . . ]]<CR/LF^EOI>
where chnum is the channel number for the channel whose output
switch is set to ON.
If no output switches are ON, the query response is:
CL<CR/LF^EOI>
1 to 10: Returns the source status of SMU or MFCMU.
The type parameter corresponds to slot number where the module is
installed.
For the SMU when the output switch is ON, the query response is:
DV chnum,range,voltage[,Icomp[,comp polarity[,irange]]]
<CR/LF^EOI>
or
DI chnum,range,current[,Vcomp[,comp polarity[,vrange]]]
<CR/LF^EOI>
where range is the present setting of the output range.
For the SMU when the output switch is OFF, the query response is:
CL chnum<CR/LF^EOI>
For the MFCMU, the query response is:
DCV chnum,voltage;ACV chnum,voltage;FC chnum,frequency
<CR/LF^EOI>