TTC
The *TST? slot,0 command just returns the pass/fail result of the latest
*TST?/CA/*CAL? command or the auto calibration.
The *TST? 0,0 command returns the memorized latest pass/fail result of all
modules.
If slot specifies the slot that installs no module, this command causes an error.
Remarks If a SMU connected to SCUU fails this command, the SCUU cannot be controlled.
And the SSP and SSL commands are not available.
If the MFCMU connected to SCUU fails this command, all modules connected to
SCUU cannot be used.
Failed modules are disabled, and can only be enabled by the RCV command. To
recover the modules connected to SCUU, recover MFCMU and SMU in this order.
Query Response results<CR/LF^EOI>
See Table 4-32. results returns the sum of the values corresponding to the failures.
Example
Statements
OUTPUT @B1500;"*TST?"
ENTER @B1500;A
Table 4-32 *TST? Response
TTC
The TTC command performs the high speed spot measurement by using the
MFCMU, and returns the measurement data and the time data from timer reset to the
start of measurement. The command starts a current measurement regardless of the
trigger mode (TM command) and the measurement mode (MM command).
results Description results Description
0 Passed. No failure detected. 32 Slot 6 module failed.
1 Slot 1 module failed. 64 Slot 7 module failed.
2 Slot 2 module failed. 128 Slot 8 module failed.
4 Slot 3 module failed. 256 Slot 9 module failed.
8 Slot 4 module failed. 512 Slot 10 module failed.
16 Slot 5 module failed. 1024 Mainframe failed.
Page view 522