Agilent B1500A/B1505A Programming Guide, Edition 11 4-177
Command Reference
SPM?
0: DC voltage output mode.
1: 2-level pulse output mode using pulse signal source 1.
2: 2-level pulse output mode using pulse signal source 2.
3: 3-level pulse output mode using pulse signal source 1 and 2.
Example
Statements
OUTPUT @B1500;"SPM 101,3"
SPM?
This query command returns the output mode of the specified SPGU channel.
Syntax SPM? chnum
Parameters chnum : SPGU channel number. Integer expression. 1 to 10 or 101 to 1002. See
Table 4-1.
Query Response mode<CR/LF^EOI>
0: DC voltage output mode.
1: 2-level pulse output mode using pulse signal source 1.
2: 2-level pulse output mode using pulse signal source 2.
3: 3-level pulse output mode using pulse signal source 1 and 2.
Example
Statements
OUTPUT @B1500;"SPM? 101"
ENTER @B1500;A
SPP
This command stops all channel outputs and all trigger outputs of the SPGU
simultaneously. The output goes to the following value.
• base value set by the SPV command, in PG mode
• initial value of waveform, in ALWG mode
This command does not stop the SPGU DC bias output.
Syntax SPP
Example
Statements
OUTPUT @B1500;"SPP"