Agilent B1500A/B1505A Programming Guide, Edition 11 4-177
Command Reference
SPM?
0: DC voltage output mode.
1: 2-level pulse output mode using pulse signal source 1.
2: 2-level pulse output mode using pulse signal source 2.
3: 3-level pulse output mode using pulse signal source 1 and 2.
Example
Statements
OUTPUT @B1500;"SPM 101,3"
SPM?
This query command returns the output mode of the specified SPGU channel.
Syntax SPM? chnum
Parameters chnum : SPGU channel number. Integer expression. 1 to 10 or 101 to 1002. See
Table 4-1.
Query Response mode<CR/LF^EOI>
0: DC voltage output mode.
1: 2-level pulse output mode using pulse signal source 1.
2: 2-level pulse output mode using pulse signal source 2.
3: 3-level pulse output mode using pulse signal source 1 and 2.
Example
Statements
OUTPUT @B1500;"SPM? 101"
ENTER @B1500;A
SPP
This command stops all channel outputs and all trigger outputs of the SPGU
simultaneously. The output goes to the following value.
base value set by the SPV command, in PG mode
initial value of waveform, in ALWG mode
This command does not stop the SPGU DC bias output.
Syntax SPP
Example
Statements
OUTPUT @B1500;"SPP"