4-204 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
TSC
Execution
Conditions
The CN/CNX command has been executed for the specified channel.
If you want to apply DC voltage over ± 25 V, the SCUU must be connected
correctly. The SCUU can be used with the MFCMU and two SMUs (MPSMU or
HRSMU). The SCUU cannot be used if the HPSMU is connected to the SCUU or if
the number of SMUs connected to the SCUU is only one.
If the output voltage is greater than the allowable voltage for the interlock open
condition, the interlock circuit must be shorted.
Syntax TMDCV chnum,mode[,range]
Parameters chnum : MFCMU channel number. Integer expression. 1 to 10 or 101 to 1001.
See Table 4-1 on page 4-16.
mode : Ranging mode. Integer expression. 0 or 2.
0: Auto ranging. Initial setting.
2: Fixed range.
range : Measurement range. This parameter must be set if mode=2. See Table
4-20 on page 4-31.
Example
Statements
OUTPUT @B1500;"TMDCV 9,0"
ENTER @B1500 USING "#,3X,13D,X";Dcbias
TSC
The TSC command enables or disables the time stamp function.
Execution
Conditions
Time stamp function is not available for the following measurement modes:
Quasi-pulsed spot measurement (MM 9)
Linear search measurement (MM 14)
Binary search measurement (MM 15)
This command is not effective for the 4 byte binary data output format (FMT3 and
FMT4).
Syntax TSC mode