4-130 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
MCPT
For current source (mode=2): See Table 4-6 on page 4-23, Table 4-8
on page 4-25, Table 4-11 on page 4-27, or Table 4-13 on page 4-28
for each measurement resource type. base and peak must have the
same polarity.
comp : Compliance (in A or V). Numeric expression. If you do not set comp,
the previous value is used.
For voltage source (mode=1): See Table 4-7 on page 4-24, Table
4-9 on page 4-26, Table 4-10 on page 4-26, Table 4-12 on page
4-27, Table 4-14 on page 4-28, or Table 4-15 on page 4-28 for each
measurement resource type.
For current source (mode=2): See Table 4-6 on page 4-23, Table 4-8
on page 4-25, Table 4-11 on page 4-27, or Table 4-13 on page 4-28
for each measurement resource type.
Compliance polarity is automatically set to the same polarity as the
output value, regardless of the specified comp. If the output value is 0,
the polarity is set to positive.
Remarks The N value and the chnum value set to the MCPNX, MCPWNX, and WNX
commands must be unique for each command execution. If you set the value used to
the previous command, the previous command setting is cleared, and the last
command setting is effective.
Source channels set by the WNX commands start output in the order specified by
the N value, and then the source channels set by the MCPNX and MCPWNX
commands start output simultaneously.
If you use multiple measurement channels, all measurement channels start
measurement simultaneously.
Example
Statements
OUTPUT @B1500;"AIT 2,3,1E-3"
OUTPUT @B1500;"MCPT 1,5E-2,2E-2,1"
OUTPUT @B1500;"MCPNT 3,1E-2,2E-2"
OUTPUT @B1500;"MCPNX 1,3,1,0,0,5,1E-1"
MCPT
The MCPT command sets the hold time, pulse period, measurement timing, and
number of measurements, which are the parameters for the multi channel pulsed
spot or sweep measurement set by MM 27 or MM 28.
Syntax MCPT hold[,period[,Mdelay[,average]]]
Page view 445