1-50 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Basics
Data Output Format
C Range. One byte. Range value used to calculate the data.
C
for SMU data for CMU data
VI C Z ACDCF
00000000 (0) 1 Ω
00000001 (1) 10 Ω
00000010 (2) 100 Ω
00000011 (3) 1 kΩ 8 V 1 kHz
00000100 (4) 10 kΩ 16 mV 12 V 10 kHz
00000101 (5) 100 kΩ 32 mV 25 V 100 kHz
00000110 (6) 1 MΩ 64 mV 1 MHz
00000111 (7) 10 MΩ 125 mV 100 V
00001000 (8) 0.5 V 1 pA 1 pF 100 MΩ 250 mV
00001001 (9) 5 V 10 pA 10 pF 1 GΩ
00001010 (10) 0.2 V 100 pA 100 pF 10 GΩ
00001011 (11) 2 V 1 nA 1 nF 100 GΩ
00001100 (12) 20 V 10 nA 10 nF
00001101 (13) 40 V 100 nA 100 nF
00001110 (14) 100 V 1 μA1 μF
00001111 (15) 200 V 10 μA10 μF
00010000 (16) 500 V 100 μA100 μF
00010001 (17) 1500 V 1 mA 1 mF
00010010 (18) 3000 V 10 mA 10 mF
00010011 (19) 10 kV 100 mA 100 mF
00010100 (20) 1 A 1 F
00010101 (21) 2 A
00010110 (22) 20 A
00010111 (23) 40 A
00011010 (26) 500 A
00011100 (28) 2000 A
00011111 (31) Invalid data is returned.