4-224 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
WNX
For the linear sweep, the B1500 uses the minimum range that
covers both start and stop values to force the staircase sweep
current.
For the log sweep, the B1500 changes the output range
dynamically.
For the limited auto ranging, the instrument never uses the range
less than the specified range.
start, stop : Start or stop value (in V or A). Numeric expression. Setting start=stop
sets the SMU to a constant source. For the log sweep, start and stop
must have the same polarity.
For voltage source (mode=1): See Table 4-7 on page 4-24, Table
4-9 on page 4-26, Table 4-10 on page 4-26, Table 4-12 on page
4-27, Table 4-14 on page 4-28, or Table 4-15 on page 4-28 for each
measurement resource type.
For current source (mode=2): See Table 4-6 on page 4-23, Table 4-8
on page 4-25, Table 4-11 on page 4-27, or Table 4-13 on page 4-28
for each measurement resource type.
comp : Compliance (in A or V). Numeric expression. If you do not set comp,
the previous value is used.
For voltage source (mode=1): See Table 4-7 on page 4-24, Table
4-9 on page 4-26, Table 4-10 on page 4-26, Table 4-12 on page
4-27, Table 4-14 on page 4-28, or Table 4-15 on page 4-28 for each
measurement resource type.
For current source (mode=2): See Table 4-6 on page 4-23, Table 4-8
on page 4-25, Table 4-11 on page 4-27, or Table 4-13 on page 4-28
for each measurement resource type.
For the log sweep and without Pcomp, set the value available for
the minimum range that covers start and stop values.
Compliance polarity is automatically set to the same polarity as the
output value, regardless of the specified comp. If the output value is 0,
the polarity is set to positive.
If you set Pcomp, the maximum comp value for the module is allowed,
regardless of the output range setting.
Pcomp : Power compliance (in W). Numeric expression. Resolution: 0.001 W. If
the Pcomp value is not entered, the power compliance is not set. This
parameter is not available for HVSMU.