Agilent B1500A/B1505A Programming Guide, Edition 11 4-231
Command Reference
WTACV
0 to delay, with 0.1 ms resolution. Numeric expression.
If this parameter is not set, Tdelay will be 0.
Mdelay : Step measurement trigger delay time (in seconds) that is the wait time
after receiving a start step measurement trigger and before starting a
step measurement.
0 to 65.535, with 0.1 ms resolution. Numeric expression.
If this parameter is not set, Mdelay will be 0.
Example
Statements
OUTPUT @B1500;"WT 5,0.1,0.1,0.1,0.1"
OUTPUT @B1500;"WT 5,0.2"
WTACV
This command sets the hold time, delay time, and step delay time for the CV (AC
level) sweep measurement (MM23). This command is also used to set the step
source trigger delay time effective for the step output setup completion trigger and
the step measurement trigger delay time effective for the start step measurement
trigger. For the trigger function, refer to “Trigger Function” on page 2-74. If you do
not enter this command, all parameters are set to 0.
Syntax WTACV hold,delay[,Sdelay[,Tdelay[,Mdelay]]]
Parameters hold : Hold time (in seconds) that is the wait time after starting measurement
and before starting delay time for the first step.
0 (initial setting) to 655.35, with 10 ms resolution. Numeric
expression.
delay : Delay time (in seconds) that is the wait time after starting to force a step
output and before starting a step measurement.
0 (initial setting) to 655.35, with 0.1 ms resolution. Numeric
expression.
Sdelay : Step delay time (in seconds) that is the wait time after starting a step
measurement and before starting to force the next step output.
0 (initial setting) to 1, with 0.1 ms resolution. Numeric expression. If
this parameter is not set, Sdelay will be 0. If Sdelay is shorter than the
measurement time, the B1500 waits until the measurement completes,
then forces the next step output.