Agilent B1500A/B1505A Programming Guide, Edition 11 2-85
Remote Mode Functions
Trigger Function
Figure 2-50 Trigger Input/Output Example, Staircase Sweep, Negative Logic
The B1500 sets the measurement conditions, sets the trigger ports, and waits for a
Start Measurement trigger.
By the Start Measurement trigger, the B1500 starts the staircase sweep
measurement.
By the Start Step Output Setup trigger, the B1500 waits until the source trigger delay
elapses, and sends the Step Output Setup Completion trigger. If the trigger is
received during the hold time, the B1500 performs this after the hold time.
By the Start Step Measurement trigger, the B1500 waits until the measurement
trigger delay elapses, executes a step measurement, and sends the Step Measurement
Completion trigger. If the trigger is received during the delay time, the B1500
performs this after the delay time.
By the next Start Step Output Setup trigger, the B1500 changes the source output
value, and waits until the source trigger delay elapses, and sends the Step Output
Setup Completion trigger. If the trigger is received during the step delay time, the
B1500 performs this after the step delay time.
After the staircase sweep measurement, the B1500 sends the Step Measurement
Completion trigger and the Measurement Completion trigger, and puts the
measurement data in the data output buffer.
: Measurement
Start Step Measurement
Start Step Output Setup (Case1)
Start Measurement
Step Measurement Completion
Step Output Setup Completion
Measurement Completion
Hold time
Delay time or more
Step delay time or more
Source trigger delay
Measurement trigger delay