5-8 Agilent B1500A/B1505A Programming Guide, Edition 11
Error Messages
Operation Error
230 Pulse source must be set.
To perform the pulsed spot measurement (MM3), send the PV or PI
command to set the pulse source.
231 Compliance must be set correctly.
Compliance was not set or an incorrect compliance value was set in the
PV, PI, PWV, or PWI command. Set the correct compliance value
effective for the pulse output.
232 Invalid pulse output setup
Check the pulse output setup and set the correct value.
233 Invalid pulse timing setup
Check the pulse timing parameters and set the correct values.
238 Too large pulse width (max. 2 s).
The maximum value of the pulse width is 2 s. And the available value
depends on the pulse period value. See “PT” on page 4-152.
239 Pulse width must be 0.01 ms or more.
Set the pulse width to 0.01 ms or more. See “PT” on page 4-152.
240 Enter QSV to set QSCV sweep source.
The QSV command must be entered to set the sweep source of the
quasi-static CV measurement.
241 QSR range value must be -9 to -14.
The range values available for the QSR command are -9 to -14. Set
one of the available values.
242 QSCV measurement was aborted.
Quasi-static CV measurement was aborted by the automatic abort
function.
243 Enter MM13 before QSZ.
Before the QSZ command, the MM13 command must be entered to set
the measurement mode.
244 Set a longer integration time to QST.
The integration time is too short to perform the offset measurement. Set
a longer value to the integration time parameter of the QST command.