4-88 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
ERHVS
0: Normal mode, initial setting
1: HVMCU DC output mode
2: Capacitance charge mode
Example
Statements
OUTPUT @B1500;"ERHVPV 2"
ERHVS
This command sets the connection status for the series resistor on the HVSMU path
of Agilent N1266A HVSMU current expander.
Execution
Conditions
Digital I/O port must be set to the N1266A control mode using the ERMOD 8
command.
The ERHVP 0 command has been already executed.
The CL command has been already executed for the SMU modules specified in the
ERHVCA command.
Syntax ERHVS state
Parameters state : Connection status of the series resistor on the HVSMU path. Integer
expression. 0 or 1.
0: Disable the series resistor connection, initial setting
1: Enable the series resistor connection.
Example
Statements
OUTPUT @B1500;"ERHVS 1"
ERHVS?
This query command returns the connection status for the series resistor on the
HVSMU path of Agilent N1266A HVSMU current expander.
Syntax ERHVS?
Query Response state<CR/LF^EOI>
Connection status of the series resistor on the HVSMU path. 0 (disconnect) or 1
(connect).
Example
Statements
OUTPUT @B1500;"ERHVS?"
ENTER @B1500;hvRHVMC