4-164 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
QSV
Parameters chnum : SMU source channel number. Integer expression. 1 to 10 or 101 to
1001. See Table 4-1 on page 4-16.
mode : Sweep mode. Integer expression. 1 or 3.
1: Linear sweep (single stair, start to stop.)
3: Linear sweep (double stair, start to stop to start.)
vrange : Ranging type. Integer expression. The output range will be set to the
minimum range that covers both start and stop values. For the limited
auto ranging, the instrument never uses the range less than the specified
range. See Table 4-4 on page 4-21.
start, stop : Start or stop voltage (in V). Numeric expression. See Table 4-7 on page
4-24.
0 to ± 100 for MPSMU/HRSMU, or 0 to ± 200 for HPSMU
This value is the upper or lower limit of the sweep output voltage.
cvoltage : Capacitance measurement voltage (in V).
The minimum value is double the resolution of the output range, and
the maximum value is 10 V. The value must be |sweep step voltage|
you desire. If you set the value greater than |sweep step voltage|, the
cvoltage is automatically set to the same value as |sweep step voltage|.
See step.
NOTE In the QSCV measurement, the B1500 executes the capacitance measurement at the
sweep steps except for the sweep start voltage and stop voltage. At each sweep step,
the capacitance measurement is executed over the voltage range: output voltage ±
cvoltage/2 (V).
step : The number of steps for the voltage sweep. 1 to 1001. Integer
expression. Define the sweep step voltage at first, then calculate the
step value by using the following formula.
step = |startstop| / |sweep step voltage| 1
If you set step=1 and |stopstart| 10, the B1500 executes a one-point
capacitance measurement between the start and stop values. Then
cvoltage value is ignored.
Icomp : Current compliance (in A). Numeric expression. See Table4-6 on page
4-14. If you do not set Icomp, the previous value is used.
Page view 479