4-140 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
MSP
Output Data The B1500 returns the data measured before any abort condition is detected.
Dummy data 199.999E+99 will be returned for the data after abort.
Example
Statements
OUTPUT @B1500;"MSC 2"
OUTPUT @B1500;"MSC 2,2"
MSP
The MSP command specifies the SPGU channel synchronized with the sampling
measurements (MM10), and the channel output after the sampling measurement.
You can also specify the channel output before the sampling measurement for the
channel which is set to the DC bias output by the SPM command.
Using the measurement trigger (XE command, etc.), the specified channel’s output
changes to the base value (set by the SPV command) and then changes to the pulse
peak value or the DC bias value (set by the SPV command) after the base hold time
elapses (which is set by the MT command). Finally the output is changed to the post
value (set by the MSP command) after the sampling measurement.
Execution
Conditions
Sampling interval (interval of MT command) must be 2 ms.
Syntax MSP chnum[,post[,base]]
Parameters chnum : SPGU channel number. Integer expression. 1 to 10 or 101 to 1002. See
Table 4-1 on page 4-16.
post : Output voltage after completing sampling measurement. Numeric
expression. If this parameter is not set, the SPGU channel outputs the
base value set by the SPV command.
0 to ± 40 V, in 1 mV resolution
base : Output voltage before starting sampling measurement. Numeric
expression. This parameter is available only for the SPGU channel
which is set to the DC bias output by the SPM command. If this
parameter is not set, the SPGU channel outputs the base value set by
the SPV command.
0 to ± 40 V, in 1 mV resolution
Remarks The MSC command’s post parameter is not effective for the SPGU channel
synchronized with the sampling measurements (MM10).