Agilent B1500A/B1505A Programming Guide, Edition 11 4-111
Command Reference
LIM
This command ignores the RV command setting.
Syntax LGV chnum,mode,range,target
Parameters chnum : SMU search monitor channel number. Integer expression. 1 to 10 or
101 to 1001. See Table 4-1 on page 4-16.
mode : Search mode. Integer expression.
0 : If the measured value target, it is the search result data.
1 : If the measured value target, it is the search result data.
range : Measurement ranging type. Integer expression. The measurement range
will be set to the minimum range that covers the target value. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See Table 4-2 on page 4-17.
target : Search target voltage (in V). Numeric expression.
0 to ±100 V (MPSMU/HRSMU)
0 to ±200 V (HPSMU)
0 to ±30 V (MCSMU)
0 to ±40 V (HCSMU/DHCSMU)
0 to ±3000 V (HVSMU)
Example
Statements
OUTPUT @B1500;"LGV 1,2,12,3"
LIM
The LIM command sets the maximum output limit of the voltage or current applied
by SMU. The setting value is effective for all of the SMU and memorized until the
LIM command is executed again. The value is not initialized by the *RST command
or power off.
Syntax LIM mode,limit
Parameters mode : Limit setup mode. 1 or 2. Integer expression.
1: Voltage output limit
2: Current output limit
limit : Maximum output limit value, in V or A. Numeric expression.
For voltage limit: 200 V to 10000 V (initial value), 100 V resolution.
For current limit: 1 A to 1500 A (initial value), in 1 A resolution.