Agilent B1500A/B1505A Programming Guide, Edition 11 4-85
Command Reference
ERHPS?
ERHPS?
The ERHPS? command returns the connection status for the series resistor on the
HVSMU path of Agilent N1258A/N1259A module selector.
Syntax ERHPS?
Query Response onoff<CR/LF^EOI>
Connection status of the series resistor on the HVSMU path. 1 (connect) or 0
(disconnect). Integer expression.
Example
Statements
OUTPUT @B1500;"ERHPS?"
ENTER @B1500;A
ERHVCA
This command specifies the modules connected to the V Control, I Control, and
HVSMU inputs of Agilent N1266A HVSMU current expander.
Execution
Conditions
No SMU may be in the high voltage state (forcing more than the allowable voltage
for the interlock open condition, or voltage compliance set to more than it).
The CL command has been already executed for the SMU modules specified in the
ERHVCA command.
Digital I/O port must be set to the N1266A control mode using the ERMOD 8
command.
Syntax ERHVCA vsmu,ismu
,hvsmu
Parameters vsmu : Slot number for the MC/HCSMU connected to the V Control input.
ismu : Slot number for the MC/HCSMU connected to the I Control input.
hvsmu : Slot number for the HVSMU connected to the HVSMU input.
Effective slot numbers are 1 to 10. Integer expression. Use the smaller slot number
for the module which occupies two slots, such as HVSMU and HCSMU.
Enter 0 to cancel the assignment.
Example
Statements
OUTPUT @B1500;"ERHVCA 3,4,9"