4-218 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
WM
Vcomp : Voltage compliance (in V). Numeric expression. See Table 4-6 on page
4-23, Table 4-8 on page 4-25, or Table 4-11 on page 4-27 for each
measurement resource type.
If you do not set Vcomp, the previous value is used. If Vcomp value is
greater than the allowable voltage for the interlock open condition, the
interlock circuit must be shorted.
Compliance polarity is automatically set to the same polarity as the
output value, regardless of the specified Vcomp. If the output value is 0,
the polarity is set to positive.
If you set Pcomp, the maximum Vcomp value for the measurement
resource is allowed, regardless of the output range setting.
For the log sweep and without Pcomp, set the value available for the
minimum range that covers start and stop values.
Pcomp : Power compliance (in W). Numeric expression. Resolution: 0.001 W. If
the Pcomp value is not entered, the power compliance is not set. This
parameter is not available for HVSMU.
0.001 to 2 for MPSMU/HRSMU, 0.001 to 20 for HPSMU, 0.001 to 40
for HCSMU, 0.001 to 80 for dual HCSMU, 0.001 to 3 for MCSMU,
0.001 to 100 for UHVU
Example
Statements
OUTPUT @B1500;"WI 1,1,11,0,0.1,100,10,1"
OUTPUT @B1500;"WI 2,2,15,1E-6,0.1,100"
WM
The WM command enables or disables the automatic abort function for the staircase
sweep sources and the pulsed sweep source. The automatic abort function stops the
measurement when one of the following conditions occurs:
Compliance on the measurement channel
Compliance on the non-measurement channel
Overflow on the AD converter
Oscillation on any channel
This command also sets the post measurement condition for the sweep sources.
After the measurement is normally completed, the staircase sweep sources force the
value specified by the post parameter, and the pulsed sweep source forces the pulse
base value.