2-24 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Measurement Modes
Sampling Measurements
Sampling measurement is performed as shown below. The sampling operation is
performed in the specified time interval until the number of measurement result data
reaches to the specified number of samples.
Figure 2-16 Sampling Measurements
1. The sampling condition is set by the MT, ML, and MSC commands.
2. The SMU synchronous source channels are set by the MV/MI commands. The
channels will start output by a trigger, such as the XE command.
3. The SPGU synchronous source channels are set by the SIM 0, SPRM, SPPER,
SPM, SPT, SPV, and MSP. The channels will start output by a trigger.
4. The source channels set by the DI/DV commands will start output at the timing
of the DI/DV command execution.
5. The synchronous source channels force the base value by a trigger such as the
XE command, as follows.
The SPGUs start output in the order from lower to higher slot number, and then
the SMUs start output simultaneously.
Voltage or current
DV/DI
MT/ML/MSC /MV/MI
Measurement trigger (e.g. XE)
Measurement time
Channel 1 output
Channel 2 output
Setup value
Previous value
Bias value
Previous value
Voltage or current
Time
: Measurement
Base hold time
Bias hold time
Base value
Sampling interval
Peak value
Previous value
Voltage
Base value
SIM0/SPRM/SPPER/
SPM/SPT /SPV/MSP
SMU
SMU
SPGU