1-36 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Basics
Data Output Format
Binary Data Output Format
This section describes the binary data output format, and the elements of the data.
•“Time Stamp
“Data Resolution”
•“Data Format
“4 Bytes Data Elements”
“8 Bytes Data Elements”
Time Stamp
The B1500 can record the time when the measurement is started, and sends the time
data (Time). This function is enabled by the TSC command. The time data will be
sent just before the measurement data.
For example, in the staircase sweep measurements, the data will be as shown below.
Block1 [Block2] . . . . <terminator>
where, BlockN (N: integer) = Time1 Data1 [Time2 Data2] ... [Source_data], then
TimeN (N: integer) is the time from timer reset to the start of DataN measurement.
The timer count is cleared/reset by the TSR command (Time=0).
The time stamp function is not available for the following measurements.
4 bytes binary data format (FMT3 or FMT4)
High speed spot measurement
Quasi-pulsed spot measurement (MM9)
Linear search measurement (MM14)
Binary search measurement (MM15)