102 Chapter 7
Processing Measurement Results
Analyzing Measurement Results
90 Cent=6.0E+7
100 Span=2.0E+4
110 Model$="E"
120 !
130 OUTPUT @Hp4294a;"PRES"
140 OUTPUT @Hp4294a;"MEAS IMPH"
150 OUTPUT @Hp4294a;"CENT ";Cent
160 OUTPUT @Hp4294a;"SPAN ";Span
170 OUTPUT @Hp4294a;"SING"
180 OUTPUT @Hp4294a;"*OPC?"
190 ENTER @Hp4294a;Buff$
200 !
210 OUTPUT @Hp4294a;"TRAC A"
220 OUTPUT @Hp4294a;"AUTO"
230 OUTPUT @Hp4294a;"TRAC B"
240 OUTPUT @Hp4294a;"AUTO"
250 OUTPUT @Hp4294a;"EQUC CIR"&Model$
260 OUTPUT @Hp4294a;"CALECPARA"
270 OUTPUT @Hp4294a;"DEFECR1?"
280 ENTER @Hp4294a;R1
290 OUTPUT @Hp4294a;"DEFECC1?"
300 ENTER @Hp4294a;C1
310 OUTPUT @Hp4294a;"DEFECL1?"
320 ENTER @Hp4294a;L1
330 OUTPUT @Hp4294a;"DEFECC0?"
340 ENTER @Hp4294a;C0
350 !
360 PRINT " --- Result ---"
370 PRINT "R1 :";R1;"[ohm]"
380 PRINT "C1 :";C1;"[F]"
390 PRINT "L1 :";L1;"[H]"
400 PRINT "C0 :";C0;"[F]"
410 !
420 OUTPUT @Hp4294a;"SIMFCHAR"
430 END
Statistics analysis
The Agilent 4294A allows you to analyze statistics (average, standard deviation, difference
between maximum and minimum). Use the following command to turn this function ON or
OFF.
“MEASTAT” on page 349
NOTE The statistics analysis function is available when the marker function is ON.
The applicable range for statistics analysis is the same as the range searched by the marker.
You can execute the above command to display analysis results. Use the following
command to read analysis results.
“OUTPMSTA?” on page 388