288 Chapter 16
GPIB Command Reference
DCALOPEN{G|C}
DCALOPEN{G|C}
Syntax DCALOPENG <numeric>[SIE]
DCALOPENC <numeric>
DCALOPEN{G|C}?
Description For user calibration data measurement (“CAL{A|B|C}” command on page 274), sets the
conductance value (“DCALOPENG” command) and the capacitance value
(“DCALOPENC” command) of the OPEN standard of the calibration kit used to measure
OPEN calibration data.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[CAL] - USER CAL -DEFINE VAUE - {OPEN CONDUCT (G) | OPEN CAP (C)}
<numeric>
Description Conductance value or capacitance value
Range
*1
-1E6 to 1E6
Initial value 0
Unit For conductance value: S (siemens)
For capacitance value: fF (femto farad)
Resolution For conductance value: 1E-12
For capacitance value: 1
*1.The unit of range value is S (siemens) or F (farad).