Application Sample Programs
Measuring Dielectric Material
1. Chapter Title
2. Chapter Title 13. Application Sample
4. Chapter Title
5. Chapter Title
Programs
1190 PAUSE
1200 DISP "Wait until the setup is finished."
1210 OUTPUT @Agt4294a;";E4TP OFF;E4TP M1"
1220 Result=FNUser_Corr(@Agt4294a,"Adapter_Phase")
1230 IF Result<>0 THEN Prog_end
1240 OUTPUT @Agt4294a;";ECALDON"
1250 DISP ""
1260 PRINT TABXY(2,5);"x] "
1270 RETURN
1280 !
1290 Adjust:!
1300 IF Elec$="C" OR Elec$="D" THEN
1310 PRINT TABXY(2,6);"n] "
1320 RETURN
1330 END IF
1340 IF F_adapt_setup=0 THEN
1350 DISP "Perform the electrode adjustment? 0:Yes, 1:Skip";
1360 INPUT "",Ans
1370 IF Ans=1 THEN
1380 PRINT TABXY(2,6);"s] "
1390 RETURN
1400 END IF
1410 END IF
1420 F_adjust=1
1430 PRINT TABXY(5,6);"->"
1440 OUTPUT @Agt4294a;"SWPT LIN"
1450 OUTPUT @Agt4294a;";CENT ";Adj_center
1460 OUTPUT @Agt4294a;";SPAN ";0
1470 OUTPUT @Agt4294a;";POIN ";2
1480 OUTPUT @Agt4294a;";;CALP USER"
1490 OUTPUT @Agt4294a;";MEAS CPD;TRAC A"
1500 OUTPUT @Agt4294a;"HIDI ON"
1510 OUTPUT @Agt4294a;"EDITLIML"
1520 OUTPUT @Agt4294a;";LIMCLEL ;LIMSADD"
1530 OUTPUT @Agt4294a;";LIMSTAR ";Adj_center
1540 OUTPUT @Agt4294a;";LIMSTOP ";Adj_center
1550 OUTPUT @Agt4294a;";LIMLSTAR ";Adj_low_limit
1560 OUTPUT @Agt4294a;";LIMUSTAR ";Adj_up_limit
1570 OUTPUT @Agt4294a;";LIMLSTOP ";Adj_low_limit
1580 OUTPUT @Agt4294a;";LIMUSTOP ";Adj_up_limit
1590 OUTPUT @Agt4294a;";LIMSDON;;LIMEDONE"
1600 OUTPUT @Agt4294a;";TOPV ";1.5*Adj_up_limit-.5*Adj_low_limit
1610 OUTPUT @Agt4294a;";BOTV ";1.5*Adj_low_limit-.5*Adj_up_limit
1620 !
1630 OUTPUT @Agt4294a;"LIMITEST OFF"
1640 DISP "Set the micrometer of 16451B at 10um and press 'Continue'."
1650 PAUSE
1660 OUTPUT @Agt4294a;"LIMITEST ON"
1670 OUTPUT @Agt4294a;"CONT"
1680 OUTPUT @Agt4294a;";MKR ON"
1690 DISP "Adjust the electrode until PASS is displayed."
1700 PAUSE
1710 OUTPUT @Agt4294a;"HOLD"
1720 OUTPUT @Agt4294a;"OUTPFAIP?"
1730 ENTER @Agt4294a;F_nop
1740 IF F_nop<>0 THEN Adj
1750 OUTPUT @Agt4294a;"LIMITEST OFF"
1760 PRINT TABXY(2,6);"x] "
1770 RETURN
1780 !
1790 Compen:!
1800 IF F_adjust=0 THEN
1810 DISP "Perform the fixture compensation? 0:Yes, 1:Skip";
1820 INPUT "",Ans