128 Chapter 9
Communication with External Equipment (Using the I/O Ports)
Using the I/O Ports
To read out data from each input port (C to E) to the controller, use the following
commands.
To use ports C, D, and E as input ports, use the following commands to set ports C and D as
input ports in advance.
“CIN” on page 278
“DIN” on page 297
To select positive logic or negative logic for input/output signals of ports, use the following
commands. You can save this setting into an instrument setup file using the save function.
“NEGL” on page 362
“POSL” on page 403
NOTE The above commands, used to change the positive logic/negative logic setting, are
available for the following ports.
Output ports A to H
Input ports C to D
PASS/FAIL signal
To make the setting so that OUTPUT1 and OUTPUT2 goes HIGH (or LOW) when a pulse
is inputted into INPUT1, use the following commands.
“OUT1ENV{H|L}” on page 367
“OUT2ENV{H|L}” on page 368
To set OUTPUT1 and OUTPUT2 to HIGH (or LOW), use the following commands.
“OUT1{H|L}” on page 367
“OUT2{H|L}” on page 368
To check that any pulses have been inputted to INPUT1, use the following command.
“INPT?” on page 319
GPIB command Instrument BASIC command Description
“OUTPINPCIO?” on page 381 READIO(16,2)” Reads out 4-bit width data from port C.
“OUTPINPDIO?” on page 382 “READIO(16,3)” Reads out 4-bit width data from port D.
“OUTPINPEIO?” on page 382 “READIO(16,4)” Reads out 8-bit width data from port E.