286 Chapter 16
GPIB Command Reference
DATOVAL
DATOVAL
Syntax DATOVAL <numeric>
DATOVAL?
Description For the active trace (set with the “TRAC” command on page 457), if the measurement
parameter is a scalar value, sets a value that is subtracted from the data trace (offset value).
The contents of the data trace is the result of subtracting the offset value from the measured
data (if the contents of the data trace is an operation result depending on the setting with the
“DISP” command on page 301, the operation result).
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Display] - OFFSET
<numeric>
Description Offset value
Range -100E6 to 100E6
Initial value 0
Resolution 1E-18