Chapter 16 357
GPIB Command Reference
MKRP
16. GPIB
Command Reference
( F - N )
MKRP
Syntax MKRP <numeric>
MKRP?
Description For the active trace (set with the “TRAC” command on page 457), moves the marker to the
specified measurement point. In the coupled marker mode (set with the “MKRCOUP”
command on page 352), regardless of the active trace setting, moves the markers on both
the A and B traces. If this command is executed as Query, it reads out the measurement
point number at the marker position. If the marker function is off, executing this command
causes an error and the command is ignored.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
In the continuous marker mode, the number of the largest measurement point that does not
exceed the marker position is reads out.
Corresponding
key
No front panel key is available to execute this function.
MKRPKD
Syntax MKRPKD
Description For the active trace (set with the “TRAC” command on page 457), sets ΔX and ΔY of the
peak definition so that the current marker position is recognized as a peak. In other words,
ΔX and ΔY are respectively set sweep parameter value difference and measurement
parameter value difference between the maker and its left adjacent measurement point.
Note that, if the Δ maker mode is on, ΔX and ΔY are respectively set sweep parameter
value difference and measurement parameter value difference between the maker and the
Δ
maker.
If the marker function is off, executing this command causes an error and the command is
ignored. (No query)
Corresponding
key
[Search] - PEAK - PEAK DEF MENU - MKR PEAK DELTA
<numeric>
Description Number of the measurement point to which the marker is moved
(the number assigned in order from the left edge assuming that the
number of the leftmost measurement point is 1)
Range 1 to the number of points
Resolution 1