Agilent Technologies 4294A Service Manual Page 350

  • Download
  • Add to my manuals
  • Print
  • Page
    / 518
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 349
350 Chapter 16
GPIB Command Reference
MINDCV
MINDCV
Syntax MINDCV <numeric>[V]
MINDCV?
Description Sets the voltage limit minimum value when the dc bias output mode is the constant-voltage
mode or constant-current mode (CVOLT or CCURR has been specified with the
“DCMOD” command on page 291).
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Source] - BIAS MENU - MIX LIMIT VOLTAGE
MKR
Syntax MKR {ON|OFF|1|0}
MKR?
Description For the active trace (set with the “TRAC” command on page 457), toggles on and off the
marker function. In the coupled marker mode (set with the “MKRCOUP” command on
page 352), the setting is applied to both the A and B traces regardless of the active trace
setting. Turning off the marker function using this command initializes the marker related
settings.
Parameters
Query response {1|0}<newline><^END>
Corresponding
key
[Marker] - MKR ON off
<numeric>
Description Voltage limit minimum value of the dc bias output
Range -40 to 40
Initial value -40
Unit V (volt)
Resolution 1E-3
Description
ON or 1 Turns on the marker function.
OFF or 0 (initial value) Turns off the marker function.