378 Chapter 16
GPIB Command Reference
OUTPDTRC?
OUTPDTRC?
Syntax OUTPDTRC?
Description Reads out the values of all measurement points in a data trace array (refer to “Internal data
arrays” on page 81). (Query only)
Query response {numeric 1},{numeric 2},..,{numeric NOP×2-1},{numeric NOP×2}<newline><^END>
Reads out the readout and subsidiary readout of the measurement parameter value of each
measurement point
as shown below. Where, NOP is the number of points, and n is an
integer between 1 and NOP.
{numeric n×2-1}: Readout of the n-th measurement point. If the measurement parameter
is a scalar value (for other than COMPLEX Z-Y), the measurement
parameter value is read out. If the measurement parameter is a vector
value (for COMPLEX Z-Y), resistance (for trace A) or conductance
(for trace B) is read out.
{numeric n×2}: Subsidiary readout of the n-th measurement point. If the measurement
parameter is a scalar value (for other than COMPLEX Z-Y), 0 is
always read out. If the measurement parameter is a vector value (for
COMPLEX Z-Y), reactance (for trace A) or susceptance (for trace B)
is reads out.
Corresponding
key
No front panel key is available to execute this function.