Agilent Technologies 4294A Service Manual Page 450

  • Download
  • Add to my manuals
  • Print
  • Page
    / 518
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 449
450 Chapter 16
GPIB Command Reference
TARL?
TARL?
Syntax TARL? <numeric 1>
Description Within the waveform analysis range specified with the “ANARANG” command on page
267, searches the analysis range from right to left for a point of the specified measurement
value, and reads out the sweep parameter value at the first detected point. If the specified
measurement parameter value is not detected, 0 is read out.(Query only)
Parameters
If the specified parameter is out of the setting range, an error occurs and the command is
ignored.
Query response {numeric 2}<newline><^END>
Corresponding
key
No front panel key is available to execute this function.
<numeric 1>
Description Measurement parameter value you want to search for
Range -9.9E37 to 9.9E37
Unit Varies depending on the measurement parameter (refer to the
explanation of unit in
“SCAL” on page 423).
{numeric 2}
Description Sweep parameter value of the detected point
Unit Varies depending on the sweep parameter (refer to the explanation
of unit in
“CENT” on page 277).
Page view 449
1 2 ... 445 446 447 448 449 450 451 452 453 454 455 ... 517 518

Comments to this Manuals

No comments