Chapter 16 459
GPIB Command Reference
TRGS
16. GPIB
Command Reference
( S - Z , :PROG )
TRGS
Syntax TRGS {INT|EXT|BUS|MAN}
TRGS?
Description Selects a trigger source.
Parameters
Query response {INT|EXT|BUS|MAN}<newline><^END>
Corresponding
key
[Trigger] - SOURCE [ ] -{FREE RUN | EXTERNAL | HP-IB/LAN | MANUAL}
USKEY
Syntax USKEY
Description Displays user keys (ON KEY LABELS) of Instrument BAISC. (No query)
Corresponding
key
[System] - IBASIC - ON KEY LABELS
Description
INT (initial value) Specifies the internal trigger.
EXT Specifies the external trigger inputted from the EXT TRIGGER
terminal on the rear panel.
BUS Specifies the GPIB/LAN trigger (trigger by executing the “*TRG”
command on page 261).
MAN Specifies the manual trigger (trigger by the following key sequence
on the front panel:
[Trigger] - SOURCE [ ] - MANUAL).