304 Chapter 16
GPIB Command Reference
DMKRP
DMKRP
Syntax DMKRP <numeric>
DMKRP?
Description For the active trace (set with the “TRAC” command on page 457), moves the Δ marker to
the specified measurement point. In the coupled marker mode (set with the “MKRCOUP”
command on page 352), moves the Δ markers of both the A and B traces, regardless of the
active trace setting. If this command is executed as Query, it reads out the measurement
point number at the Δ marker position. If the Δ marker mode is off, executing this
command causes an error and the command is ignored.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
No front panel key is available to execute this function.
<numeric>
Description Number of the destination measurement point (number given in
order from the left edge assuming that the number of the left edge
measurement point is 1)
Range 1 to the number of points
Initial value 1
Resolution 1