Chapter 16 433
GPIB Command Reference
SEATARG
16. GPIB
Command Reference
( S - Z , :PROG )
SEATARG
Syntax SEATARG <numeric>[DEG|RAD|OHM|SIE|H|F|PCT]
SEATARG?
Description For the active trace (set with the “TRAC” command on page 457), sets the target value
(value in the Y axis direction) when performing the target search in the marker search
function. In the Δ marker mode, the value set with this command is dealt as a relative value
form the Δ marker. Therefore, to set the target value, use a relative value from the Δ
marker.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Search] - TARGET - TARGET VALUE
SEATARGL
Syntax SEATARGL {ON|OFF|1|0}
SEATARGL?
Description For the active trace (set with the “TRAC” command on page 457), enables/disables the
display of the line indicating the target value (target line) when performing the target
search in the marker search function.
Parameters
Query response {1|0}<newline><^END>
Corresponding
key
[Search] - TARGET - LINE on OFF
<numeric>
Description Target value at the execution of the target search
Range -100M to 100M
Initial value 0
Unit Varies depending on the measurement parameter (refer to the
explanation of unit in
“SCAL” on page 423).
Description
ON or 1 Enables the display of the target line.
OFF or 0
(initial value)
Disables the display of the target line.