386 Chapter 16
GPIB Command Reference
OUTPMIN?
OUTPMIN?
Syntax OUTPMIN?
Description Reads out the value of the minimum measurement parameter value within the waveform
analysis range set with the
“ANARANG” command on page 267 and the sweep parameter
value at the position. If there are several measurement points of the minimum measurement
parameter value, reads out the value of the leftmost measurement point within the analysis
range.(Query only)
Query response {numeric 1},{numeric 2}<newline><^END>
Readouts are as follows:
{numeric 1}: Minimum measurement parameter value.
{numeric 2}: Sweep parameter value at the position of the minimum measurement
parameter value
Corresponding
key
No front panel key is available to execute this function.
OUTPMINMAX?
Syntax OUTPMINMAX?
Description Within the waveform analysis range set with the “ANARANG” command on page 267,
reads out the maximum measurement parameter value and the sweep parameter value at
the position and the minimum measurement parameter value and the sweep parameter
value at the position. If there are several measurement points of the minimum or maximum
measurement parameter value, reads out the value of the leftmost measurement point
within the analysis range. (Query only)
Query response {numeric 1},{numeric 2},{numeric 3},{numeric 4}<newline><^END>
Readouts are as follows:
{numeric 1}: Minimum measurement parameter value.
{numeric 2}: Sweep parameter value at the position of the minimum measurement
parameter value.
{numeric 3}: Maximum measurement parameter value.
{numeric 4}: Sweep parameter value at the position of the maximum measurement
parameter value.
Corresponding
key
No front panel key is available to execute this function.