274 Chapter 16
GPIB Command Reference
BWFACT
BWFACT
Syntax BWFACT {1|2|3|4|5}
BWFACT?
Description Sets the bandwidth. To set the bandwidth of each segment when creating the list sweep
table, also use this command.
Parameters
Note that, if you use this command to create an additional segment, the setting of the
previous segment is used as its initial value.
Query response {numeric}<newline><^END>
Corresponding
key
[Bw/Avg] - BAND WIDTH [ ] - {1 FAST | 2 | 3 | 4 | 5 PRECISE}
CAL{A|B|C}
Syntax CAL{A|B|C}
Description Measures OPEN/SHORT/LOAD data for user calibration. The “CALA” command
measures OPEN data; the “CALB” command, SHORT data; the “CALC” command,
LOAD data. (No query)
Corresponding
key
[CAL] - USER CAL - EXCUTE CAL - {OPEN[ ] | SHORT[ ] | LOAD[ ]}
CALDON
Syntax CALDON
Description Finishes the measurement of user calibration data, calculates the calibration coefficient
from the measurement result, and stores it into the volatile memory (RAM). Executing this
command automatically turns on the user calibration function (specified to ON with the
“CALST” command on page 276). If all the measurements of OPEN/SHORT/LOAD data
are not completed, executing this command causes an error and the command is ignored.
(No query)
Corresponding
key
[CAL] - USER CAL - EXCUTE CAL - done
Description
1 (Initial value) Specifies bandwidth 1 (shortest measurement time).
2 Specifies bandwidth 2.
3 Specifies bandwidth 3.
4 Specifies bandwidth 4.
5 Specifies bandwidth 5 (longest measurement time, accurate
measurement).