GPIB Command Reference
FILC
16. GPIB
Command Reference
( F - N )
FILC
Syntax FILC <string 1>,<string 2>,<string 3>,<string 4>
Description Copies a file. To specify a file, use a file name including its extension (refer to Table 8-1 on
page 113). (No query)
Parameters
Corresponding
key
[Save] - FILE UTILITIES - COPY FILE
FMT
Syntax FMT {LINY|LOGY|POLA|COMP}
FMT?
Description For the active trace (set with the “TRAC” command on page 457), selects the format to
display the trace. Depending on the setting of the measurement parameter, the settable
display format differs. Trying to select a non-settable display format causes an error and
the command is ignored.
Parameters For scalar measurement parameters:
For vector measurement parameters:
Query response {LINY|LOGY|POLA|COMP}<newline><^END>
Corresponding
key
[Format] - {LIN | LOG | COMPLEX | POLAR}
<string 1> <string 2> <string 3> <string 4>
Description Copy source
file name
Copy source
device name
Copy
destination
file name
Copy destination
device name
Range 12 characters
or less
(including the
extension)
Select from
"MEMO",
"DISK", or
"FLASH"
*1
12 characters
or less
(including the
extension)
Select from
"MEMO", "DISK",
or "FLASH"
*1
*1.Each selection item indicates the device name as shown below.
"MEMO": RAM disk (volatile)
"DISK": Diskette
"FLASH": Flash disk (non-volatile)
Description
LINY (initial value) Specifies the linear Y axis format.
LOGY Specifies the log Y axis format.
Description
POLA Specifies the polar chart format.
COMP (initial value) Specifies the complex plane format.