Chapter 16 429
GPIB Command Reference
SEARMAX
16. GPIB
Command Reference
( S - Z , :PROG )
SEARMAX
Syntax SEARMAX <numeric>[HZ|MHZ|V|A]
SEARMAX?
Description For the marker search function for the active trace (set with the “TRAC” command on page
457), if the partial search is enabled (specified to ON with the “PARS” command on page
395), sets the partial search range upper border value.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Search] - SEARCH RANGE MENU - MAXIMUM
<numeric>
Description Upper border value of the partial search range
When the sweep
parameter is
frequency
Range 40 to 110E6
Initial value 110E6
Unit Hz
Resolution 1E-3
When the sweep
parameter is
OSC level (voltage)
Range 5E-3 to 1
Initial value 1
Unit V (volt)
Resolution 1E-3
When the sweep
parameter is
OSC level (current)
Range 200E-6 to 20E-3
Initial value 20E-3
Unit A (ampere)
Resolution 1E-6
When the sweep
parameter is
dc bias level (voltage)
Range -40 to 40
Initial value 40
Unit V (volt)
Resolution 1E-3
When the sweep
parameter is
dc bias level (current)
Range -100E-3 to 100E-3
Initial value 100E-3
Unit A (ampere)
Resolution 20E-6