GPIB Command Reference
OUTPSMKR{1-7}?
16. GPIB
Command Reference
( O - R )
Figure 16-6 Analyzed parameters
OUTPSMKR{1-7}?
Syntax OUTPSMKR{1|2|3|4|5|6|7}?
Description For the active trace (set with the “TRAC” command on page 457), reads out the
measurement parameter value and sweep parameter value at the sub marker (marker
number 1 to 7) position. If the sub marker is off, executing this command causes an error
and the readouts are invalid. (Query only)
Query response {numeric 1},{numeric 2},{numeric 3}<newline><^END>
Readouts are as follows:
{numeric 1}: Readout of the measurement parameter value at the sub marker
position. If the measurement parameter is a scalar value (for other than
COMPLEX Z-Y), the measurement parameter value is read out. If the
measurement parameter is a vector value (for COMPLEX Z-Y), the
readout is as follows.
If the polar chart format is selected, the amplitude is read out.
If the complex plane format is selected, resistance (for trace A) or
conductance (for trace B) is read out.
{numeric 2}: Subsidiary readout of the measurement parameter value at the sub
marker position. If the measurement parameter is a scalar value (for
other than COMPLEX Z-Y), 0 is always read out. If the measurement
parameter is a vector value (for COMPLEX Z-Y), the readout is as
follows:
If the polar chart format is selected, the phase is read out.
If the complex plane format is selected, reactance (for trace A) or
susceptance (for trace B) is read out.
{numeric 3}: Sweep parameter value at the sub marker position.
Corresponding
key
No front panel key is available to execute this function.