Agilent Technologies 4294A Service Manual Page 333

  • Download
  • Add to my manuals
  • Print
  • Page
    / 518
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 332
Chapter 16 333
GPIB Command Reference
LIMSEGM
16. GPIB
Command Reference
( F - N )
LIMSEGM
Syntax LIMSEGM <numeric>
LIMSEGM?
Description Sets the execution target segment used when no segment is specified at the execution of the
“LIMSEDI” command on page 332, “LIMSADD” command on page 330, and
“LIMSDEL” command on page 331. The setting of the execution target segment is
initialized when the creation/edit of the limit line table is finished.
If the creation/edit of the limit line table is not started, executing this command
automatically starts table creation/edit and then sets the execution target segment.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
If the creation/edit of the limit line table is not started, 0 is always read out.
Corresponding
key
[System] - LIMIT TEST - EDIT LIMIT LINE - SEGMENT
LIMSQUI
Syntax LIMSQUI
Description In the creation/edit of the limit line table, aborts segment creation/edit. (No query)
Corresponding
key
[System] - LIMIT TEST - EDIT LIMIT LINE - cancel
<numeric>
Description Segment number you want to specify as the edit target
Range 1 to the number of the segments in the limit line table
Initial value Segment number of the last line in the limit line table
Resolution 1
Page view 332
1 2 ... 328 329 330 331 332 333 334 335 336 337 338 ... 517 518