Agilent Technologies 4294A Service Manual Page 125

  • Download
  • Add to my manuals
  • Print
  • Page
    / 518
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 124
Chapter 9 125
Communication with External Equipment (Using the I/O Ports)
Using the I/O Ports
1. Chapter Title
2. Chapter Title 3. Chapter Title
9. Communication with
5. Chapter Title
Exernal Equipment
(Using the I/O Ports)
Table 9-1 Signal assignment
Pin number Signal name Signal specification
1 GND 0 V
2 INPUT1TTL TTL level, pulse input (width: 1 μs or more)
3 OUTPUT1 TTL level, latch output
4 OUTPUT2 TTL level, latch output
5 Output port A0 TTL level, latch output
6 Output port A1 TTL level, latch output
7 Output port A2 TTL level, latch output
8 Output port A3 TTL level, latch output
9 Output port A4 TTL level, latch output
10 Output port A5 TTL level, latch output
11 Output port A6 TTL level, latch output
12 Output port A7 TTL level, latch output
13 Output port B0 TTL level, latch output
14 Output port B1 TTL level, latch output
15 Output port B2 TTL level, latch output
16 Output port B3 TTL level, latch output
17 Output port B4 TTL level, latch output
18 Output port B5 TTL level, latch output
19 Output port B6 TTL level, latch output
20 Output port B7 TTL level, latch output
21 Input/output port C0 TTL level, latch output
22 Input/output port C1 TTL level, latch output
23 Input/output port C2 TTL level, latch output
24 Input/output port C3 TTL level, latch output
25 Input/output port D0 TTL level, latch output
26 Input/output port D1 TTL level, latch output
27 Input/output port D2 TTL level, latch output
28 Input/output port D3 TTL level, latch output
29 Port C status TTL level (input mode: LOW, output mode: HIGH)
30 Port D status TTL level (input mode: LOW, output mode: HIGH)
31 Write strobe signal TTL level, negative logic, pulse output (width: 10 μs; typical)
32 +5V pull-up
33 SWEEP END signal TTL level, negative logic, pulse output (width: 7 μs; typical)
34 +5V +5V, 100 mA MAX
35 PASS/FAIL signal TTL level (PASS: HIGH, FAIL: LOW), latch output
36 PASS/FAIL write strobe
signal
TTL level, negative logic, pulse output (width: 10 μs; typical)
Page view 124