258 Chapter 16
GPIB Command Reference
*CLS
IEEE common command
This section describes the IEEE common commands.
*CLS
Syntax *CLS
Description Clears the error queue, Status Byte Register, Operation Status Register, Standard Event
Status Register, and Instrument Event Status Register. This command has the same
function as the
“CLES” command on page 279. (No query)
Corresponding
key
No front panel key is available to execute this function.
*ESE
Syntax *ESE <numeric>
*ESE?
Description Sets the value of the Standard Event Status Enable Register.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
No front panel key is available to execute this function.
<numeric>
Description Setting value of the register
Range 0 to 255
Initial value 0
Resolution 1