394 Chapter 16
GPIB Command Reference
OUTPSWPRM?
OUTPSWPRM?
Syntax OUTPSWPRM?
Description Reads out the sweep parameter values of all measurement points. (Query only)
Query response {numeric 1},{numeric 2},..,{numeric NOP}<newline><^END>
Where, NOP is the number of points.
Corresponding
key
No front panel key is available to execute this function.
OUTPSWPRMP?
Syntax OUTPSWPRMP? <numeric 1>
Description Reads out the sweep parameter value of the specified measurement point. (Query only)
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric 2}<newline><^END>
Corresponding
key
No front panel key is available to execute this function.
OUTPVAC?
Syntax OUTPVAC?
Description When the oscillator level monitor function is set to on (set to ON with “OMON” command
on page 365), reads out the oscillator voltage level monitor values of all measurement
points. (Query only)
Query response {numeric 1},{numeric 2},..,{numeric NOP}<newline><^END>
Where, NOP is the number of points.
If the oscillator level monitor function is set to off, the readouts are invalid.
Corresponding
key
No front panel key is available to execute this function.
<numeric 1>
Description Number of the measurement point you want to read out (the
number assigned in order from the left edge assuming that the
number of the leftmost measurement point is 1)
Range 1 to the number of points
Resolution 1
{numeric 2}
Description Read-out sweep parameter value